Conductivity-to-TDS Conversion Values
Calibration Standard Bottles
| Model #. |
Conductivity value |
TDS value: ppm KCl |
TDS value: ppm NaCl |
TDS value: ppm 442 |
| WD-00653-23 |
23 µS |
11.6 |
10.7 |
14.74 |
| WD-00653-16 |
84 µS |
40.38 |
38.04 |
50.50 |
| WD-00653-47 |
447 µS |
225.6 |
215.5 |
300.0 |
| WD-00653-18 |
1413 µS |
744.7 |
702.1 |
1000 |
| WD-00653-15 |
1500 µS |
757.1 |
737.1 |
1050 |
| WD-00653-27 |
2070 µS |
1045 |
1041 |
1500 |
| WD-00653-20 |
2764 µS |
1382 |
1414.8 |
2062.7 |
| WD-00653-89 |
8974 µS |
5101 |
4487 |
7608 |
| WD-00606-10 |
12,880 µS |
7447 |
7230 |
11,367 |
| WD-00653-50 |
15,000 µS |
8759 |
8532 |
13,455 |
| WD-00653-32 |
80 mS |
52,168 |
48,384 |
79,688 |
Calibration Standard Pouches
| Model #. |
Conductivity value |
TDS value: ppm KCl |
TDS value: ppm NaCl |
TDS value: ppm 442 |
| WD-35653-09 |
10µS |
4.7 |
4.8 |
7.0 |
| WD-35653-10 |
447 µS |
225.6 |
215.5 |
300.0 |
| WD-35653-11 |
1413 µS |
744.7 |
702.1 |
1000 |
| WD-35653-12 |
2764 µS |
1382 |
1414.8 |
2062.7 |
| WD-35653-13 |
15,000 µS |
8759 |
8532 |
13,455 |
go back
Oakton Instruments
P.O. Box 5136, Vernon Hills, IL 60061
info@4oakton.com
|